کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8159859 1524994 2018 21 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Referenced MR thermometry using three-echo phase-based fat water separation method
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم فیزیک ماده چگال
پیش نمایش صفحه اول مقاله
Referenced MR thermometry using three-echo phase-based fat water separation method
چکیده انگلیسی
A three-point image reconstruction method for internally referenced MR thermometry was developed. The technique exploits the fact that temperature-induced changes in the water resonance frequency are small relative to the chemical shift difference between water and fat signals. This property enabled the use of small angle approximations to derive an analytic phase-based fat-water separation method for MR thermometry. Ethylene glycol and cream cool-down experiments were performed to validate measurement technique. Over a cool-down temperature range of 20 °C, maximum deviation between probe and MR measurement (averaged over 1.3 cm3 region surrounding probe) was 0.6 °C and 1.1 °C for ethylene glycol and cream samples, respectively.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Magnetic Resonance Imaging - Volume 49, June 2018, Pages 86-93