کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8170150 1526303 2016 13 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
First tests of the applicability of γ-ray imaging for background discrimination in time-of-flight neutron capture measurements
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
First tests of the applicability of γ-ray imaging for background discrimination in time-of-flight neutron capture measurements
چکیده انگلیسی
In this work we explore for the first time the applicability of using γ-ray imaging in neutron capture measurements to identify and suppress spatially localized background. For this aim, a pinhole gamma camera is assembled, tested and characterized in terms of energy and spatial performance. It consists of a monolithic CeBr3 scintillating crystal coupled to a position-sensitive photomultiplier and readout through an integrated circuit AMIC2GR. The pinhole collimator is a massive carven block of lead. A series of dedicated measurements with calibrated sources and with a neutron beam incident on a 197Au sample have been carried out at n_TOF, achieving an enhancement of a factor of two in the signal-to-background ratio when selecting only those events coming from the direction of the sample.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 823, 1 July 2016, Pages 107-119
نویسندگان
, , , , , , , , , , ,