کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8174234 1526351 2015 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Implementation of autocorrelation method for investigations of the scintillator decay time and space correlation of radiation
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Implementation of autocorrelation method for investigations of the scintillator decay time and space correlation of radiation
چکیده انگلیسی
A new method to determine scintillator decay time and space correlation of radiation is described. The method is based on an autocorrelation delayed-coincidence time spectrometer and it has been applied to measure the decay time of inorganic and organic scintillators in a range from tens of nanoseconds to a few microseconds. Results with NAFTALEN and CsI(Tl) show decay times of τ=75.0(8) ns and 998(15) ns, respectively.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 775, 1 March 2015, Pages 148-153
نویسندگان
, , ,