کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8175983 1526363 2014 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Neutron-sensitive ZnS/10B2O3 ceramic scintillator detector as an alternative to a 3He-gas-based detector for a plutonium canister assay system
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Neutron-sensitive ZnS/10B2O3 ceramic scintillator detector as an alternative to a 3He-gas-based detector for a plutonium canister assay system
چکیده انگلیسی
A neutron-sensitive ZnS/10B2O3 ceramic scintillator detector was developed as an alternative to a 3He-gas-based detector for use in a plutonium canister assay system. The detector has a modular structure, with a flat ZnS/10B2O3 ceramic scintillator strip that is installed diagonally inside a light-reflecting aluminium case with a square cross-section, and where the scintillation light is detected using two photomultiplier tubes attached at both ends of the case. The prototype detectors, which have a neutron-sensitive area of 30 mm×250 mm, exhibited a sensitivity of 21.7-23.4±0.1 cps/nv (mean±SD) for thermal neutrons, a 137Cs gamma-ray sensitivity of 1.1-1.9±0.2×10−7 and a count variation of less than 6% over the detector length. A trial experiment revealed a temperature coefficient of less than −0.24±0.05%/°C over the temperature range of 20-50 °C. The detector design and the experimental results are presented.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 763, 1 November 2014, Pages 340-346
نویسندگان
, , , , , , , , , , , , , , ,