Examining the departure in response of non-point detectors due to non-uniform illumination and displacement of effective center
ترجمه فارسی عنوان
بررسی خروج در پاسخ از آشکارسازهای غیر نقطه به دلیل روشنایی غیر یکنواخت و جابجایی مرکز موثر
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
چکیده انگلیسی
A mathematical simulation approach based on the general purpose Monte Carlo N-particle transport code MCNP was developed to calculate the departure in reading of the neutron spectrometer instrument from that expected according to the inverse square law. The calculations were performed to evaluate the effects of beam divergence on the response of a 10 in. spherical device equipped with a long BF3 counter irradiated by 11 mono-energy neutron beams. The necessary geometry correction factor, because of non-uniform illumination, for the calibration of seven polyethylene spheres with several radionuclide neutron sources, i.e. Ra-Be, 241Am-Be, 241Am-B and Po-Be sources was also determined. In all calculations, the displacement of effective center from the geometric center of moderating spheres, when used as an instrument for neutron fluence measurement, was quantified.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 728, 11 November 2013, Pages 145-149