کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8179205 1526394 2013 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Beam test of novel n-in-p strip sensors for high radiation environment
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Beam test of novel n-in-p strip sensors for high radiation environment
چکیده انگلیسی
Highly radiation tolerant n-in-p strip sensors have been developed for the high-luminosity LHC (HL-LHC). This paper reports the results of measurements with 392 MeV kinetic energy proton beam at RCNP in December 2011. The data was taken with a new DAQ system consisting of an universal read-out board 'SEABAS' and beam tracking telescopes whose spacial resolution is better than 5μm. The aim of this beam test is to evaluate the new 1 cm×1 cm n-in-p miniature sensors before and after 1015 neq cm−2 irradiation. The median charge of un-irradiated sensor is 6.2 fC at full depletion voltage, while the median charge after 1015 neq cm−2 irradiation of the sensor is 4.2 fC. The novel Punch-Through-Protection (PTP) has been implemented in these sensors. The length of active region of the sensor around PTP is observed to be decreased by 12μm in the irradiated sensors at 1015 neq cm−2.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 731, 11 December 2013, Pages 247-251
نویسندگان
, , , , , , , , , , , , , , , , , ,