کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8179372 1526399 2013 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Spatial resolution of a μPIC-based neutron imaging detector
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Spatial resolution of a μPIC-based neutron imaging detector
چکیده انگلیسی
We present a detailed study of the spatial resolution of our time-resolved neutron imaging detector utilizing a new neutron position reconstruction method that improves both spatial resolution and event reconstruction efficiency. Our prototype detector system, employing a micro-pattern gaseous detector known as the micro-pixel chamber (μPIC) coupled with a field-programmable-gate-array-based data acquisition system, combines 100μm-level spatial and sub-μs time resolutions with excellent gamma rejection and high data rates, making it well suited for applications in neutron radiography at high-intensity, pulsed neutron sources. From data taken at the Materials and Life Science Experimental Facility within the Japan Proton Accelerator Research Complex (J-PARC), the spatial resolution was found to be approximately Gaussian with a sigma of 103.48±0.77μm (after correcting for beam divergence). This is a significant improvement over that achievable with our previous reconstruction method (334±13μm), and compares well with conventional neutron imaging detectors and with other high-rate detectors currently under development. Further, a detector simulation indicates that a spatial resolution of less than 60μm may be possible with optimization of the gas characteristics and μPIC structure. We also present an example of imaging combined with neutron resonance absorption spectroscopy.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 726, 21 October 2013, Pages 155-161
نویسندگان
, , , , , , , , , , , , , , , , , , ,