کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8179449 1526396 2013 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Single channel beta-gamma coincidence system for radioxenon measurement using well-type HPGe and plastic scintillator detectors
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Single channel beta-gamma coincidence system for radioxenon measurement using well-type HPGe and plastic scintillator detectors
چکیده انگلیسی
In order to improve detection sensitivity for radioxenon isotopes, a new single channel beta-gamma coincidence system has been developed. The system combines a well-type High-purity Germanium (HPGe) detector to measure gamma or X radiation and a plastic scintillator detector to obtain electron radiation. A 133Xe sample has been produced and the radioactive concentration was determined with length-compensated method based on three internal gas proportional counters. The performance of system has been checked by measuring 133Xe sample with and without coincidence. The coincidence detection efficiency of 81 keV gamma-ray from decay of 133Xe was calibrated to be 0.34 (1±2.4%), and the Minimum detectable activity (MDA) of the beta-gamma coincidence system for 133Xe was determined to be 1.8 mBq after one day of measurement.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 729, 21 November 2013, Pages 900-904