کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8215999 1533017 2015 23 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Single-Isocenter Multiple-Target Stereotactic Radiosurgery: Risk of Compromised Coverage
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Single-Isocenter Multiple-Target Stereotactic Radiosurgery: Risk of Compromised Coverage
چکیده انگلیسی
The effects of rotational errors on target coverage were studied across a broad range of SRS cases. In general, the risk of compromised coverage increased with decreasing target volume, increasing rotational error and increasing distance between targets. Multivariate regression models from this study may be used to quantify the dosimetric effects of rotational errors on target coverage given patient-specific input parameters of PTV volume and distance to isocenter.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: International Journal of Radiation Oncology*Biology*Physics - Volume 93, Issue 3, 1 November 2015, Pages 540-546
نویسندگان
, , , , ,