کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8233531 1533106 2010 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Comparison of Planned Dose Distributions Calculated by Monte Carlo and Ray-Trace Algorithms for the Treatment of Lung Tumors With CyberKnife: A Preliminary Study in 33 Patients
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Comparison of Planned Dose Distributions Calculated by Monte Carlo and Ray-Trace Algorithms for the Treatment of Lung Tumors With CyberKnife: A Preliminary Study in 33 Patients
چکیده انگلیسی
We establish that changing the treatment plan calculation algorithm from EPL to MC can produce large differences in target and critical organs' dose coverage. The observed discrepancies are larger for plans using smaller collimator sizes and have strong dependency on the anatomical relationship of target-critical structures.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: International Journal of Radiation Oncology*Biology*Physics - Volume 77, Issue 1, 1 May 2010, Pages 277-284
نویسندگان
, , , , , ,