Comparison of CT on Rails With Electronic Portal Imaging for Positioning of Prostate Cancer Patients With Implanted Fiducial Markers
چکیده انگلیسی
Considerable differences between in-room CT and EPI exist. The phantom measurements showed slice thickness affected the accuracy of localization in the SI plane, and couch sag that occurs at the CT on rails gantry could not be totally corrected for in the AP plane. Other confounding factors are the action of rotating the couch and associated time lag between image acquisitions (prostate motion), EPI image quality, and outlining uncertainties.