Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table
چکیده انگلیسی
Conclusion: The XVI image acquisition and registration procedure were highly reproducible. Both translational and rotational positioning errors can be corrected very precisely with the HRTT. The HRTT is therefore well suited to complement cone-beam computed tomography to take full advantage of position correction in six degrees of freedom for IGRT. The combination of XVI and the HRTT has the potential to improve the accuracy of high-precision treatments.