کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8253161 1533599 2018 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Experimental and Monte Carlo-based determination of the beam quality specifier for TomoTherapyHD treatment units
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
Experimental and Monte Carlo-based determination of the beam quality specifier for TomoTherapyHD treatment units
چکیده انگلیسی
The stopping power ratio water-to-air sw,aΔ directly depends on the beam quality specifier. The value determined from all experimental TPR20,10 data was sw,aΔ=1.126±0.1%, which is in excellent agreement with the value directly calculated by Monte Carlo simulations. The agreement is a good indication that the equations proposed by Sauer and Palmans are able to calculate the beam quality specifier under reference conditions from measurements in arbitrary photon field sizes with high accuracy and are applicable for the TomoTherapyHD treatment unit.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Zeitschrift für Medizinische Physik - Volume 28, Issue 2, April 2018, Pages 142-149
نویسندگان
, , , ,