کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
826948 907964 2008 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Development of a Peristaltic Micropump for Bio-Medical Applications Based on Mini LIPCA
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
Development of a Peristaltic Micropump for Bio-Medical Applications Based on Mini LIPCA
چکیده انگلیسی

This paper presents the design, fabrication, and experimental characterization of a peristaltic micropump. The micropump is composed of two layers fabricated from Polydimethylsiloxane (PDMS) material. The first layer has a rectangular channel and two valve seals. Three rectangular mini lightweight piezo-composite actuators are integrated in the second layer, and used as actuation parts. Two layers are bonded, and covered by two Polymethyl Methacrylate (PMMA) plates, which help increase the stiffness of the micropump. A maximum flow rate of 900 μL·min−1 and a maximum backpressure of 1.8 kPa are recorded when water is used as pump liquid. We measured the power consumption of the micropump. The micropump is found to be a promising candidate for bio-medical application due to its bio-compatibility, portability, bidirectionality, and simple effective design.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Bionic Engineering - Volume 5, Issue 2, June 2008, Pages 135-141