کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
850499 909286 2013 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
PZT calibration
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی (عمومی)
پیش نمایش صفحه اول مقاله
PZT calibration
چکیده انگلیسی

One important source of error in phase-shift interferometry consists of a nonuniform application of phase steps, due, for example, to a nonlinear piezoelectric effect in the phase shifter. We developed a method for inline PZT calibration using a video camera for acquiring interferograms and an ATI Radeon X1650 video card for frame processing. The PZT was calibrated over a voltage range of 40 V and a displacement range of 11 μm. We acquired 801 interferograms in a Twyman-Green interferometer and a 801 × 801 matrix computed from the subtraction of each interferogram from every other interferogram was obtained and used for calibration. The speed and accuracy of frame processing of the video card was compared to the results obtained by an Intel QuadCore processor. In terms of accuracy there are no significant differences, but the processing using the video card was ∼5 times faster than the processing using the processor. The processing took ∼5 min and we obtained a dispersion of ∼10 nm when we fitted the displacement of the PZT to a polynomial of degree 5 in voltage.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Optik - International Journal for Light and Electron Optics - Volume 124, Issue 17, September 2013, Pages 2803–2806
نویسندگان
, , , ,