کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
855061 1470697 2015 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Automated information-measurement system used to test the immunity of the medical equipment protective elements to the discharges of the defibrillator
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی (عمومی)
پیش نمایش صفحه اول مقاله
Automated information-measurement system used to test the immunity of the medical equipment protective elements to the discharges of the defibrillator
چکیده انگلیسی

The article discusses the information-measurement system used for testing of protective elements medical equipment for their resistance to the discharge of a defibrillator, which is designed to test medical products, which may be exposed to high voltage defibrillator pulses, up to 5000 V while in operation. The Information-measurement system should provide a fully automated testing process in accordance with the international standard IEC 60601-1, as well as analyze the result and diagnose the parameters of elements the of a defibrillator block simulating the pulses.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Procedia Engineering - Volume 129, 2015, Pages 239-244