کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
858168 1470739 2014 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Biofilm Oxygen Profiling using an Array of Microelectrodes on a Microfabricated Needle
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی (عمومی)
پیش نمایش صفحه اول مقاله
Biofilm Oxygen Profiling using an Array of Microelectrodes on a Microfabricated Needle
چکیده انگلیسی

A novel microelectrode array (DO-MEA) sensor was designed and fabricated using microelectromechanical systems technology on a needle for real time measurement of dissolved oxygen (DO). The sensor consisted of eleven gold disk microelectrodes and a rectangular auxiliary electrode along them. The sensor can also be operated with an integrated reference system. Three different sensor designs were fabricated, and their responses were fully characterized and evaluated. The DO-MEA sensor presented a linear response in the 0-8 mg DO·L-1 concentration range in water, displaying high sensitivity and repeatability. Knowledge of bacterial activity inside biofilms is key to the optimization of applied biotechnologies. The developed sensor was validated against a commercial Clark-type microelectrode overcoming its drawbacks, by profiling a heterotrophic biofilm cultivated in a flat-plate bioreactor. The DO-MEA sensor provided a multipoint, simultaneous dissolved oxygen snapshot profile inside a biofilm with high spatial resolution due to its micrometric dimensions, thus becoming a powerful tool for the research of many similar biological-based processes and applications.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Procedia Engineering - Volume 87, 2014, Pages 256-259