کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8669704 1578448 2018 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Balloon assisted retrieval of broken catheter in a 10-months old baby with patent ductus arteriosus: Case report
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Balloon assisted retrieval of broken catheter in a 10-months old baby with patent ductus arteriosus: Case report
چکیده انگلیسی
Percutaneous retrieval of intravascular foreign body can be done by snares, biopsy forceps, dormia basket or tip deflecting wires. Herein, we report percutaneous retrieval a fractured multipurpose diagnostic catheter by balloon assisted technique in a 10-months old baby while doing device closure of a patent ductus arteriosus. The catheter got fractured at the junction of shaft and the tip while manipulation to right ventricular outflow tract during antegrade wiring. The fractured proximal end was trapped at the tip of venous sheath which was successfully wired with percutaneous coronary angioplasty wire and retrieved with the help of 1.5 × 10 mm and 2 × 10 mm Maverick balloon (Boston scientific; USA). This is the first ever report of balloon assisted retrieval of intravascular foreign body.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Indian College of Cardiology - Volume 8, Issue 2, June 2018, Pages 95-98
نویسندگان
, , , , , ,