کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8718080 1588484 2018 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
External Defibrillator Damage Associated With Attempted Synchronized Dual-Dose Cardioversion
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی طب اورژانس
پیش نمایش صفحه اول مقاله
External Defibrillator Damage Associated With Attempted Synchronized Dual-Dose Cardioversion
چکیده انگلیسی
The simultaneous use of 2 external defibrillators to administer either dual or sequential cardioversion or defibrillation for refractory cardiac arrhythmias is increasing in both the out-of-hospital and inhospital settings. Using 2 defibrillators to administer higher energy levels than can be achieved with a single defibrillator is considered off-label and is currently not part of published advanced cardiac life support guidelines. We report the first case in which the use of dual-dose cardioversion was associated with external defibrillator damage. Because defibrillator damage, especially if undetected, jeopardizes patient safety and off-label medical product use may void the manufacturer's warranty, this case should urge users to proceed with caution when contemplating this technique.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Annals of Emergency Medicine - Volume 71, Issue 1, January 2018, Pages 109-112