Selections of the cylinder implant neck taper and implant end fillet for optimal biomechanical properties: A three-dimensional finite element analysis
چکیده انگلیسی

In this paper, effects of the implant neck taper and the implant-end fillet on the maximum Von Mises stresses were evaluated in jaw bones, and maximum displacements examined in an implant–abutment complex by a finite element method (FEM). The implant-neck tapers (T) ranged from 45° to 70°, and fillets of implant ends (R) ranged from 0.5 to 1.5 mm. Results suggested that under axial load by the maximum Von Mises stresses in cortical and cancellous bones decreased by 71.6% and 14.8%, respectively, and under 45° buccolingual load by 68.2% and 11.0%, respectively. The maximum displacement of implant–abutment complex decreased by 9.1% and 22.8% under axial and 45° buccolingual load, respectively. When T ranged from 64° to 73° and R exceeded 0.8 mm, minimum stress/displacement was obtained and the evaluating targets were more sensitive to T than to R. Data indicated that the taper of implant neck favored stress distribution in cortical bones more than the fillet of implant end did; taper of implant neck affected implant stability more than the fillet of implant end did; and the taper of implant neck of 64–73° and fillet of implant end exceeding 0.8 mm were optimal selections for the type B/2 bone in a cylinder implant by biomechanical consideration.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Biomechanics - Volume 41, Issue 5, 2008, Pages 1124–1130
نویسندگان
, , , , , , ,