Comparison of the pulse shape discrimination performance of plastic scintillators coupled to a SiPM
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Comparison of the pulse shape discrimination performance of plastic scintillators coupled to a SiPM
چکیده انگلیسی
Analysis was performed using a Synchronous Charge Integration Pulse Shape Discrimination (PSD) algorithm which was applied to data acquired from a mixed fast neutron/gamma radiation field from an AmBe neutron source. The collected pulses were processed offline with the energy and PSD parameters calculated. The quality of the PSD performance was characterised by a common figure of merit (FoM). The best n-γ separation was found by the newer Eljen EJ-276 scintillator with a FoM value of 3.03 ± 0.03 at an energy of 1.5 MeV gamma equivalent. The Amcrys UPS-113NG material achieved a FoM value of 2.60 ± 0.04.