Position sensitivity of the TIGRESS 32-fold segmented HPGe clover detector
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Position sensitivity of the TIGRESS 32-fold segmented HPGe clover detector
چکیده انگلیسی
Sensitivity to the locations of γ-ray interactions within a 32-fold segmented clover-type HPGe detector has been investigated through 90∘ Compton scattering of 661.6keVγ-rays from a collimated Cs137 source. A mean position sensitivity of 0.44 mm at an energy of 373 keV is deduced by comparing the average pulse shapes for net charge collecting signals, as well as transient induced signals in neighbouring contacts, from 1007 pairs of three-dimensionally localized interaction points. The reconstruction of individual event locations based on a χ2 comparison with the measured set of basis waveforms is presented.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 540, Issues 2–3, 21 March 2005, Pages 348-360