کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
9845818 1526521 2005 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Pulse shaping for Ge-spectrometers optimized for ballistic deficit and electronic noise
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Pulse shaping for Ge-spectrometers optimized for ballistic deficit and electronic noise
چکیده انگلیسی
For large-volume high-purity Ge detectors, working at low counting rates, a new two-level shaping is proposed which is based on a cusp-like form of the bottom part of the shaping and a parabolic-like form of the top shaping part. Due to the side wings of the cusp the good noise characteristics of the shaping are conserved. At the same time, the parabolic part of the pulse shape allows rather satisfactory compensation of the ballistic deficit. Calculation shows that the noise factor can be improved within 10-12% as compared to standard quasi-Gaussian shaping.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 538, Issues 1–3, 11 February 2005, Pages 718-722