کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10081735 1604093 2005 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Premature Failure of the Polyethylene Tibial Bearing Surface of the Interax Knee Arthroplasty
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Premature Failure of the Polyethylene Tibial Bearing Surface of the Interax Knee Arthroplasty
چکیده انگلیسی
Thirty-four patients (37 knees) were treated with the Interax knee arthroplasty. There were 31 female and 3 male patients with a mean age at surgery of 70 years (range, 53-84 years). Premature failure of the polyethylene tibial bearing surface occurred in 9 patients (9 knees). Eight patients of these required a revision operation because of early wearing out of the tibial surface 20 to 38 months after the operation. The failure was due to inadequate implant design in small component and polyethylene gamma irradiated in the air.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 20, Issue 5, August 2005, Pages 658-663