کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10081818 1604096 2005 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Polyethylene Failure of the Patellar Component in New Jersey Low-Contact Stress Total Knee Arthroplasties
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Polyethylene Failure of the Patellar Component in New Jersey Low-Contact Stress Total Knee Arthroplasties
چکیده انگلیسی
In a postoperative 5- to 12-year follow-up study of 598 New Jersey low-contact stress total knee arthroplasties, 32 required revision. All retrieved patellar components were examined for polyethylene damage. These 32 cases had 3 types of failure: split rupture (7 cases), peripheral wear (21), and cantilevering breakage (4), respectively 75%, 64.3%, and 100% of which showed subluxation and/or tilting of the patellar component on the prerevision roentgenograms. Misalignment at the joint contact surfaces and rotational blockage of the mobile patellar component were considered the major causes of the failure. A design of a flatter metallic button (giving larger focal thickness of the polyethylene) and a dome-shaped polyethylene (reducing stress concentration at the pinnacle) may alleviate the failure driving mechanism should subluxation or tilting of the patellar component take place.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 20, Issue 2, February 2005, Pages 202-208
نویسندگان
, , , , , ,