کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10081934 1604098 2005 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Early Femoral Head Penetration of a Highly Cross-Linked Polyethylene Liner vs a Conventional Polyethylene Liner
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Early Femoral Head Penetration of a Highly Cross-Linked Polyethylene Liner vs a Conventional Polyethylene Liner
چکیده انگلیسی
Forty highly cross-linked, polyethylene-liner primary total hip arthroplasties were matched by sex, age, and body mass index with 40 conventional polyethylene primary total hip arthroplasties to compare femoral head penetration rates. Both inserts were 10° liners manufactured by the same company and packaged in a nitrogen/vacuum atmosphere. Identical acetabular components and a 28-mm cobalt chrome femoral head were used in all cases. The mean femoral head penetration rate for the highly cross-linked and conventional polyethylene was 0.05 (range, 0.01-0.09) and 0.12 (range, 0.02-0.29) mm/y, respectively, a 58.33% reduction in femoral head penetration (P < .001). In this short-term follow-up study (mean, 47.73 months), both reduced “bedding-in” and reduced wear may contribute to the observed decrease in femoral head penetration.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 20, Supplement 3, October 2005, Pages 73-76
نویسندگان
, , ,