Analysis and evaluation of a biomedical polycarbonate urethane tested in an in vitro study and an ovine arthroplasty model. Part II: in vivo investigation
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی بیو مهندسی (مهندسی زیستی)
پیش نمایش صفحه اول مقاله
Analysis and evaluation of a biomedical polycarbonate urethane tested in an in vitro study and an ovine arthroplasty model. Part II: in vivo investigation
چکیده انگلیسی
The polyurethane (PU) elastomer Corethane 80A (Corvita) is being considered as the acetabular bearing material in a novel total replacement hip joint. Its biostability was investigated in vitro (Analysis and evaluation of a biomedical polycarbonate urethane tested in an in vitro study and an ovine arthroplasty model. Part I: material selection and evaluation, Biomaterials, in press) together with three other commercially available biomedical PUs: Pellethane 2363-80A (DOW Chemical), a polyhexamethylene oxide based PU, PHMO-PU (CSIRO, not supplied as a commercial product) and ChronoFlex AL-80A (CardioTech). From the in vitro studies, Corethane 80A displayed the best overall resistance to hydrolysis, ESC, MIO and calcification, followed by ChronoFlex 80A and PHMO-PU, with Pellethane 80A being the least stable. Building on the in vitro investigation, the follow-up in vivo study (reported here) assessed Corethane 80A as the bearing layer in a prototype compliant layer acetabular cup, in a fully functioning ovine total hip arthoplasty (THA) model. PU degradation in the retrieved cups was analysed using a range of analytical and physical-testing methods including mechanical testing, differential scanning calorimetry, Fourier transform infrared spectroscopy and environmental scanning electron microscopy. The Corethane 80A functioned well in the THA model, with the bearing surfaces of the retrieved hip cups showing no significant evidence of biodegradation or wear damage after 3 years in vivo. The findings in this study provide compelling evidence for the biostability and effectiveness of acetabular cups incorporating a Corethane 80A compliant bearing layer.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biomaterials - Volume 26, Issue 6, February 2005, Pages 633-643
نویسندگان
, , , , ,