LC-MS/MS systematic toxicological analysis: Comparison of MS/MS spectra obtained with different instruments and settings
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی زیست شیمی
پیش نمایش صفحه اول مقاله
LC-MS/MS systematic toxicological analysis: Comparison of MS/MS spectra obtained with different instruments and settings
چکیده انگلیسی
MS/MS spectra do not offer the expected inter-instrument reproducibility despite an attempt at standardizing the fragmentation conditions using a reference compound. However, although the inter-instrument differences in ion relative intensity were significant, the spectra obtained looked almost similar. This suggests that in library searching algorithms, higher weight should be assigned for the m/z ratios than for their relative intensity in the spectra.