کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10981494 1108077 2010 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Technical note: Comparing calibration methods for determination of protein in goat milk by ultraviolet spectroscopy
موضوعات مرتبط
علوم زیستی و بیوفناوری علوم کشاورزی و بیولوژیک علوم دامی و جانورشناسی
پیش نمایش صفحه اول مقاله
Technical note: Comparing calibration methods for determination of protein in goat milk by ultraviolet spectroscopy
چکیده انگلیسی
A rapid spectroscopic method to determine total protein in bovine and buffalo milk using UV spectra of guanidine-hydrochloride mixed milk has previously been reported and validated. The method was based on mixed calibration samples and univariate calibrations of fourth derivative (4D) spectra. In this study the same method was compared and tested for determination of total protein in goat milk. Calculations based on multivariate calibration (partial least squares regression) on full spectra of goat milk were used. The method was tested on 2 UV instruments. The comparison resulted in a significantly more robust (i.e., better) transferability between UV instruments for the partial least squares regression method on full spectra compared with previous univariate calibration of 4D spectra. Local (1 instrument) calibrations gave similar, significantly not different (chi-squared test) cross-validated prediction error results for the 2 methods. It can be concluded that there is no need for fourth derivation. Partial least squares regression on full spectra was equal or superior to using the 4D spectra.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Dairy Science - Volume 93, Issue 7, July 2010, Pages 2922-2925
نویسندگان
, , , , ,