کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
11032031 1645696 2018 15 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Type testing of a locally made LiF:Mg,Ti + PTFE TLD for its use as a personal dosimeter
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Type testing of a locally made LiF:Mg,Ti + PTFE TLD for its use as a personal dosimeter
چکیده انگلیسی
Our research group has developed a thermoluminescence dosimeter (TLD) based on pellets of LiF:Mg,Ti mixed with polytetrafluoroethylene (LiF:Mg,Ti +PTFE). This TLD can be used as a personal dosimeter. Extensive type testing, carried out with reference to the International Electrotechnical Commission (IEC) Standard, were performed for the purpose of accepting the LiF:Mg,Ti+PTFE as a personal TL dosimeter. Tests performed include repeatability, batch homogeneity, linearity, detection threshold, and light sensitivity. Results showed that locally made LiF:Mg,Ti+PTFE TLDs met all the standard requirements.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 141, November 2018, Pages 246-249
نویسندگان
, , ,