کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
11032034 1645696 2018 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
CT examination effective doses in Saudi Arabia
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
CT examination effective doses in Saudi Arabia
چکیده انگلیسی
Patient effective doses and the associated radiation risks arising from particular computed tomography (CT) imaging procedures are assessed. The objectives of this research are to measure radiation doses for patients and to quantify the radiogenic risks from CT brain and chest procedures. Patient data were collected from five calibrated CT modality machines in Saudi Arabia. The results are from a study of a total of 60 patients examined during CT procedures using the calibrated CT units. For CT brain and chest, the mean patient effective doses were 1.9 mSv (with a range of 0.6-2.5 mSv) and 7.4 mSv (with a range of 0.5-34.8 mSv) respectively. The radiogenic risk to patients ranged from between 10−5 and 10−4 per procedure. With 65% of the CT procedure cases diagnosed as normal, this prompts re-evaluation of the referral criteria. The establishment of diagnostic reference levels (DRL) and implementation of radiation dose optimisation measures would further help reduce doses to optimal values.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 141, November 2018, Pages 261-265