کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
11949 769 2005 10 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Electrochemical corrosion and metal ion release from Co-Cr-Mo prosthesis with titanium plasma spray coating
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی بیو مهندسی (مهندسی زیستی)
پیش نمایش صفحه اول مقاله
Electrochemical corrosion and metal ion release from Co-Cr-Mo prosthesis with titanium plasma spray coating
چکیده انگلیسی

The corrosion behavior of CoCrMo implants with rough titanium coatings, applied by different suppliers by either sintering or vacuum plasma spraying, has been evaluated and compared with uncoated material. The open-circuit potential, corrosion current and polarization resistance were determined by electrochemical techniques. The Co, Cr and Ti ions released from the samples into the electrolyte during a potentiostatic extraction technique were analyzed using ICP-MS.The Ti coatings from the different suppliers showed a different porous morphology, and the implants exhibited a distinct corrosion activity, underlining the importance of the coating process parameters. Among the titanium coated samples, the one with the sintered overcoat turned out to be the most resistant. Yet, on an absolute scale, they all showed a corrosion resistance inferior to that of uncoated CoCrMo or wrought titanium.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biomaterials - Volume 26, Issue 23, August 2005, Pages 4747–4756
نویسندگان
, , , ,