کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1228580 968479 2006 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A multi-pumping flow system for chemiluminometric determination of ascorbic acid in powdered materials for preparation of fruit juices
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A multi-pumping flow system for chemiluminometric determination of ascorbic acid in powdered materials for preparation of fruit juices
چکیده انگلیسی

A multi-pumping flow-based procedure with chemiluminescent detection is proposed for the determination of ascorbic acid, AA, in fruit juices (powdered form). The method relies on the inhibitory effect of AA on the oxidation of luminol by hydrogen peroxide in alkaline medium. The system comprises several discretely actuated solenoid pumps as the only active components. It handles 100 samples per hour, and requires 96 μl of sample, 42 μg of luminol and 105 μg of potassium hexacyanoferrate(III) per determination. The analytical curve is linear up to about 11 mmol l− 1 AA, and detection limit is 0.17 mmol l− 1 AA. The system yields precise measurements (r.s.d. < 1%; n = 11), and recovery ranges from 94% to 106%. Results are in agreement with the reference method (AOAC) at the 95% confidence level.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Microchemical Journal - Volume 83, Issue 2, July 2006, Pages 70–74