Using a monochromator to improve the resolution in TEM to below 0.5 Å. Part II: Application to focal series reconstruction
موضوعات مرتبط
مهندسی و علوم پایه مهندسی مواد فناوری نانو (نانو تکنولوژی)
پیش نمایش صفحه اول مقاله
Using a monochromator to improve the resolution in TEM to below 0.5 Å. Part II: Application to focal series reconstruction
چکیده انگلیسی

We apply monochromated illumination to improve the information transfer in focal series reconstruction to 0.5 Å at 300 kV. Contrary to single images, which can be taken arbitrarily close to Gaussian focus in a CS-corrected microscope, images in a focal series are taken at a certain defocus. This defocus poses limits on the spatial coherence of the illumination, and through this, limits on the brightness of the monochromated illumination. We derive an estimate for the minimum spatial coherence and the minimal brightness needed for a certain resolution at a certain defocus and apply this estimate to our focal series experiments. We find that the 0.5 Å information transfer would have been difficult and probably impossible to obtain without the exceptionally high brightness of the monochromated illumination.


► A monochromator is used to minimize the effect of chromatic aberration.
► We demonstrate improvement of information transfer in TEM to 0.5 Å in focal series reconstruction.
► We derive an estimate which relates the brightness of the illumination to the obtainable resolution.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ultramicroscopy - Volume 118, July 2012, Pages 35–43