کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1822294 1526324 2015 11 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A method for establishing absolute full-energy peak efficiency and its confidence interval for HPGe detectors
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
A method for establishing absolute full-energy peak efficiency and its confidence interval for HPGe detectors
چکیده انگلیسی

A method is proposed for establishing the absolute efficiency calibration of a HPGe detector including the confidence interval in the energy range of 79.6–3451.2 keV. The calibrations were accomplished with the 133Ba, 60Co, 56Co and 152Eu point-like radioactive sources with only the 60Co source being activity calibrated to an accuracy of 2% at the 90% confidence level. All data sets measured from activity calibrated and uncalibrated sources were fit simultaneously using the linearized least squares method. The proposed fit function accounts for scaling of the data taken with activity uncalibrated sources to the data taken with the high accuracy activity calibrated source. The confidence interval for the fit was found analytically using the covariance matrix. Accuracy of the fit was below 3.5% at the 90% confidence level in the 79.6–3451.2 keV energy range.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 802, 1 December 2015, Pages 102–112
نویسندگان
, , , , , ,