Study on response function of CdTe detector
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Study on response function of CdTe detector
چکیده انگلیسی

So far the origin of the mechanism of light emission in the sonoluminescence has not elucidated whether it is due to blackbody radiation or bremsstrahlung. The final goal of our study is measuring X-ray energy spectrum using high-sensitivity cadmium telluride (CdTe) detector in order to obtain information for understanding sonoluminescence phenomena. However, the scope of this report is the measurement of X-ray spectrum using a high-resolution CdTe detector and determination of CdTe detector response function to obtain the corrected spectrum from measured soft X-ray source spectrum. In general, the measured spectrum was distorted by the characteristics of CdTe detector. Monte Carlo simulation code, MCNP, was used to obtain the reference response function of the CdTe detector. The X-ray spectra of 57Co, 133Ba, and 241Am were obtained by a 4×4×1.0(t) mm3 CdTe detector at room temperature.