کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1828623 1526484 2008 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Analysis of the BC501A neutron detector signals using the true pulse shape
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Analysis of the BC501A neutron detector signals using the true pulse shape
چکیده انگلیسی

A Digital Pulse Shape Analysis method has been developed and tested for a BC501A liquid scintillator. The method is of general use and is especially suited for measurements in high counting rate environments where particle discrimination is necessary. It consists of a single-parameter fit of the true pulse shape   of the detector to the digitized signals and is based on very basic assumptions on the behavior of the detector signal. Thus, it can be extended to other detector types like inorganic scintillators and semiconductor detectors. The energy resolution, the quality of the particle discrimination and the efficiency for reconstructing pile-up events have been investigated with standard neutron and γγ-ray sources using an 8 bit and 1 GSample/s digitizer.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 597, Issues 2–3, 1 December 2008, Pages 212–218
نویسندگان
, , , , , ,