کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1829856 | 1526493 | 2007 | 5 صفحه PDF | دانلود رایگان |
عنوان انگلیسی مقاله ISI
Response of a Micro Pixel Chamber to heavy ions with the energy of several hundreds of MeV/nMeV/n
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
ابزار دقیق
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
Beam tests were performed for a Micro Pixel Chamber (μμ-PIC) with a detection volume of 10×10×10cm3 to investigate the response to heavy ions. The three-dimensional tracks of carbon, silicon, and iron beams were successfully observed and their track lengths were measured. Additionally, Linear Energy Transfer (LET) distributions of each ion were obtained, and the mean LET values were consistent with the theoretical calculation of mass stopping power within an error of ∼10%∼10%. This detector is a candidate for an ideal dosimeter in space.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 581, Issues 1–2, 21 October 2007, Pages 110–114
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 581, Issues 1–2, 21 October 2007, Pages 110–114
نویسندگان
Tsutomu Nagayoshi, Tadayoshi Doke, Yasunobu Fujita, Kaori Hattori, Koji Ishida, Jun Kikuchi, Hisashi Kitamura, Tatsuto Komiyama, Hidetoshi Kubo, Haruhisa Matsumoto, Kentaro Miuchi, Hironobu Nishimura, Kiwamu Saito, Shin-ichi Sasaki, Hiroyuki Sekiya,