کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1876177 | 1041983 | 2012 | 4 صفحه PDF | دانلود رایگان |

The extrapolation chamber is a parallel-plate ionization chamber that allows variation of its air-cavity volume. In this work, an experimental study and MCNP-4C Monte Carlo code simulations of an ionization chamber designed and constructed at the Calibration Laboratory at IPEN to be used as a secondary dosimetry standard for low-energy X-rays are reported. The results obtained were within the international recommendations, and the simulations showed that the components of the extrapolation chamber may influence its response up to 11.0%.
► A homemade extrapolation chamber was studied experimentally and with Monte Carlo.
► It was characterized as a secondary dosimetry standard, for low energy X-rays.
► Several characterization tests were performed and the results were satisfactory.
► Simulation showed that its components may influence the response up to 11.0%.
► This chamber may be used as a secondary standard at our laboratory.
Journal: Applied Radiation and Isotopes - Volume 70, Issue 7, July 2012, Pages 1388–1391