کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1876179 1041983 2012 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Quality control methodology and implementation of X-radiation standards beams, mammography level, following the standard IEC 61267
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Quality control methodology and implementation of X-radiation standards beams, mammography level, following the standard IEC 61267
چکیده انگلیسی

This study presents the results of the establishment of a quality control program developed and applied for the X-ray system of the Calibration Laboratory of IPEN. The X-ray standard beams, mammography level, using molybdenum and aluminum as additional filtration were established after the application of this quality control and the spectrometry of these qualities was made. The reference ionization chamber has traceability to the PTB. The radiation qualities RQR-M, RQA-M, RQN-M and RQB-M, following the recommendations of the IEC 61267 and the IAEA TRS 457 were established.


► Determination of the additional filtration for each mammography radiation quality.
► Determination of the KQ values for the WAV and WMV radiation qualities.
► Determination of the air-kerma rates for all radiation qualities.
► The uncertainties have been calculated using the type A and type B uncertainties.
► Establishment of the qualities that use a PMMA phantom (WAN, WAB, WMN and WMB).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 70, Issue 7, July 2012, Pages 1396–1399
نویسندگان
, , ,