کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1876240 | 1041989 | 2012 | 6 صفحه PDF | دانلود رایگان |

The yield of 478 and 558 keV gamma-rays have been measured from water samples containing 0.031–0.500 wt. % boron and 0.0625–0.500 wt. % cadmium, respectively, using a cylindrical 76 mm×76 mm (height×diameter) LaCl3:Ce detector. Inspite of interferences between detector-associated and the sample-associated prompt gamma rays, the LaCl3:Ce detector has excellent resolution for the low energy prompt gamma-rays. An excellent agreement has been observed between the experimental and calculated yield of boron and cadmium prompt gamma ray from water samples.
► Performance Tests of a large LaCl3:Ce Detector with Low Energy Prompt Gamma Rays.
► Photopeak efficiency estimation of the LaCl3:Ce detector.
► Neutron Activation spectrum of the LaCl3:Ce Detector.
► Boron and Cadmium concentration measurements in bulk water samples.
Journal: Applied Radiation and Isotopes - Volume 70, Issue 5, May 2012, Pages 882–887