کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1876283 1041994 2012 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Evaluation of uncertainty and detection limits in 210Pb and 210Po measurement in water by alpha spectrometry using 210Po spontaneous deposition onto a silver disk
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Evaluation of uncertainty and detection limits in 210Pb and 210Po measurement in water by alpha spectrometry using 210Po spontaneous deposition onto a silver disk
چکیده انگلیسی

An easy and accurate method for the determination of 210Pb and 210Po in water using 210Po spontaneous deposition onto a silver disk is proposed and assessed for its detection capabilities according to the ISO Guide for the expression of uncertainty in measurement (GUM) and ISO Standard 11929-7 concerning the evaluation of the characteristic limits for ionizing radiation measurements. The method makes no assumption on the initial values of the activity concentrations of 210Pb, 210Bi and 210Po in the sample to be analyzed, and is based on the alpha spectrometric measurement of 210Po in two different aliquots: the first one measured five weeks after the sampling date to ensure radioactive equilibrium between 210Pb and 210Bi and the second after a sufficient time for the ingrowth of 210Po from 210Pb to be significant. As shown, for a recommended time interval of seven months between 210Po measurements, the applicability of the proposed method is limited to water samples with a 226Ra to 210Pb activity ratio CRa/CPb≤4, as usual in natural waters. Using sample and background counting times of 24 h and 240 h, respectively, the detection limit of the activity concentration of each radionuclide at the sampling time for a 1 L sample typically varies between 0.7 and 16 mBq L−1 for 210Pb in water samples with an initial activity of 210Po in the range 0–200 mBq L−1, and between 0.6 and 8.5 mBq L−1 for 210Po in water samples with an initial activity of 210Pb in the same range.


► 210Pb and 210Po measurement in water by 210Po spontaneous deposition onto silver disks.
► 210Pb and 210Po determination based on 210Po measurement in two different aliquots.
► Evaluation of characteristic limits in radioactivity measurements using ISO 11929-7.
► 10Pb -210Po detection limits decrease with time elapsed between 210Po measurements.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 70, Issue 4, April 2012, Pages 758–764
نویسندگان
, , ,