کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1876305 1041998 2012 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
[18F]DPA-714, [18F]PBR111 and [18F]FEDAA1106—Selective radioligands for imaging TSPO 18 kDa with PET: Automated radiosynthesis on a TRACERLAb FX-FN synthesizer and quality controls
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
[18F]DPA-714, [18F]PBR111 and [18F]FEDAA1106—Selective radioligands for imaging TSPO 18 kDa with PET: Automated radiosynthesis on a TRACERLAb FX-FN synthesizer and quality controls
چکیده انگلیسی

Imaging of TSPO 18 kDa with PET is more and more considered as a relevant biomarker of inflammation in numerous diseases. Development of new radiotracers for TSPO 18 kDa has seen acceleration in the last years and the challenge today is to make available large amounts of such a radiotracer in compliance with GMP standards for application in humans. We present in this technical note automated productions of [18F]DPA-714, [18F]PBR111 and [18F]FEDAA1106, three promising radiotracers for TSPO 18 kDa imaging, using a TRACERlab FX-FN synthesizer. This note also includes the quality control data of the validation batches for the manufacturing qualification of clinical production of [18F]DPA-714.


► Protein TSPO 18 kDa is recognized as a biomarker of inflammation involved in many diseases.
► Radiotracers targeting TSPO prepared in compliance with GMPs are mandatory as new imaging tools.
► An automated radiosynthesis of promising radiotracers and full QC have been implemented.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 70, Issue 3, March 2012, Pages 489–497
نویسندگان
, , , , , , , , , ,