Comparison of calibration results for an extrapolation chamber obtained with different 90Sr+90Y secondary standard sources
چکیده انگلیسی

Two secondary standard systems of beta radiation were used to calibrate a PTW extrapolation chamber Model 23391. Three 90Sr+90Y sources of different activities were used in this calibration procedure. Medium-term stability of the response of the chamber was also studied. The calibration was performed with and without field-flattening filters. The relative standard deviation of the obtained calibration factors was 8.3% for the aluminum collecting electrode and 4.1% for the graphite collecting electrode.


► 90Sr+90Y standard sources were used to calibrate a PTW extrapolation chamber.
► Characterization tests of the chamber response were performed.
► Chamber response showed very good short- and medium-term stabilities.
► Linear extrapolation curves were obtained.
► Calibration factors of the chamber were acceptable.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 70, Issue 2, February 2012, Pages 388–391