کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1876363 | 1042004 | 2012 | 4 صفحه PDF | دانلود رایگان |

Two secondary standard systems of beta radiation were used to calibrate a PTW extrapolation chamber Model 23391. Three 90Sr+90Y sources of different activities were used in this calibration procedure. Medium-term stability of the response of the chamber was also studied. The calibration was performed with and without field-flattening filters. The relative standard deviation of the obtained calibration factors was 8.3% for the aluminum collecting electrode and 4.1% for the graphite collecting electrode.
► 90Sr+90Y standard sources were used to calibrate a PTW extrapolation chamber.
► Characterization tests of the chamber response were performed.
► Chamber response showed very good short- and medium-term stabilities.
► Linear extrapolation curves were obtained.
► Calibration factors of the chamber were acceptable.
Journal: Applied Radiation and Isotopes - Volume 70, Issue 2, February 2012, Pages 388–391