کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1876396 1042008 2011 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Dosimetry of gamma chamber blood irradiator using Gafchromic EBT film
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Dosimetry of gamma chamber blood irradiator using Gafchromic EBT film
چکیده انگلیسی

Gafchromic films are increasingly being used for dosimetry in medical and industrial applications of ionizing radiation because of their favorable characteristics such as self developing in nature, easy to handle, dose rate independent response, insensitivity to normal room lights, high resolution and insensitivity to variation in the environment conditions. The dosimetry measurements using Gafchromic EBT film, Fricke dosimeter and thermoluminescence (TL) dosimeter were carried out on gamma blood irradiators to establish the suitability of Gafchromic EBT film in such applications. The dose rates determined by Gafchromic EBT film, Fricke dosimeter and TL dosimeter powder are found in good agreement to one another within the uncertainty of measurement. The two-dimensional dose distributions determined using Gafchromic EBT film reflects the expected dose distribution inside the sample compartment of the blood irradiator. The results of this study establish Gafchromic EBT film a suitable dosimeter for routine dosimetry on gamma blood irradiators.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 69, Issue 1, January 2011, Pages 130–135
نویسندگان
, , , , , , ,