کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1876415 | 1042008 | 2011 | 7 صفحه PDF | دانلود رایگان |
![عکس صفحه اول مقاله: A new and simple calibration-independent method for measuring the beam energy of a cyclotron A new and simple calibration-independent method for measuring the beam energy of a cyclotron](/preview/png/1876415.png)
This work recommends a new and simple-to-perform method for measuring the beam energy of an accelerator. The proposed method requires the irradiation of two monitor foils interspaced by an energy degrader. The primary advantage of the proposed method, which makes this method unique from previous energy evaluation strategies that employ the use of monitor foils, is that this method is independent of the detector efficiency calibration. This method was evaluated by performing proton activation of natCu foils using both a cyclotron and a tandem Van de Graaff accelerator. The monitor foil activities were read using a dose calibrator set to an arbitrary calibration setting. Excellent agreement was noted between the nominal and measured proton energies.
Journal: Applied Radiation and Isotopes - Volume 69, Issue 1, January 2011, Pages 247–253