کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1876446 1042009 2011 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Performance of optically stimulated luminescence Al2O3 dosimeter for low doses of diagnostic energy X-rays
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Performance of optically stimulated luminescence Al2O3 dosimeter for low doses of diagnostic energy X-rays
چکیده انگلیسی

Personal dosimeters measure the radiation dose from exposure to hazardous sources outside the body. The present manuscript evaluates the performance of a commercially available optically stimulated luminescence (OSL) Al2O3 dosimeter using diagnostic energy X-rays. The OSL system satisfies the ANSI N13.11-2001 performance criteria for low dose diagnostic energy X-rays. Non-uniformity of sensitivity, dose linearity, X-ray energy response, and angular performance are all within the criteria of IEC-62387-1(2007).


► We evaluate the performance of a commercial optically stimulated luminescence dosimeter using low doses of diagnostic energy X-rays.
► OSL system satisfies the ANSI N13.11-2001 performance criteria for low dose diagnostic energy X-rays.
► Non-uniformity of sensitivity, dose linearity, X-ray energy response, and angular performance are all within the criteria of IEC-62387-1(2007).
► Results of this study show that OSL dosimeters may be useful for extremity dosimetry for radiation workers and patients exposed to diagnostic X-rays.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 69, Issue 10, October 2011, Pages 1486–1489
نویسندگان
, , ,