کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1876622 | 1042024 | 2010 | 6 صفحه PDF | دانلود رایگان |
In December 2003 the irradiator of the Portuguese 60Co irradiation facility, UTR, was replenished. Monte Carlo simulation studies were used to characterize the facility. Comparison of simulation results with experimental measurements were useful in estimating where the source elements were positioned and what the total activity was. For this purpose the MCNPX code was used. From the experimental point of view ionization chambers and PMMA dosemeters were used.The results indicate that the geometry of the irradiator used for simulation studies shows a shift on the vertical dimension relative to the actual geometry of approximately 8 cm, but with reduced influence on the irradiation process. Considering the uncertainties of the experimental and simulated approaches the activity of the irradiator stated by supplier has been confirmed.
Journal: Applied Radiation and Isotopes - Volume 68, Issue 1, January 2010, Pages 190–195