کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1876753 1042031 2009 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Correction factors determination in large samples gamma assay using its own multi-gamma lines spectrum
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Correction factors determination in large samples gamma assay using its own multi-gamma lines spectrum
چکیده انگلیسی

An easy and simple method for gamma assay of large multi-gamma lines samples was introduced in this work. This method performs the assay using point source calibration. The correction factors for volume and self-attenuation are experimentally deduced from the spectra of different thicknesses samples utilizing the following two simple well known facts: large and small samples of the same homogenous material have identical specific activities; the self-attenuation of gamma line decreases as its energy increases. The method was successfully applied to naturally occurring radioactive material (NORM) large samples. This method does not require complicated mathematical procedures. Neither sample matrix data nor detector unit composition is needed.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 67, Issue 10, October 2009, Pages 1819–1823
نویسندگان
, ,