The purpose of this study is to evaluate the dosimetric properties of a flattening filter free 6 MV photon beam. The 6 MV photon beam of a Varian Clinac 21EX linac was modeled using the MCNP4C Monte Carlo (MC) code. Dosimetric features including central axis absorbed doses, beam profiles and photon energy spectra were calculated for flattened and unflattened 6 MV photon beams. A substantial increase in the dose rate was seen for the unflattened beam, which was decreased with field size and depth. The penumbra width was decreased less than 0.2 mm (about 5%) and a 25% decrease in out-of-field dose was observed for the unflattened beam. The photon energy spectra were softer for the unflattened beam and the mean energies of spectra were higher for smaller field size. Our study showed that increase in the dose rate and lower out-of-field dose could be considered as practical advantages for unflattened 6 MV beams.
Journal: Applied Radiation and Isotopes - Volume 65, Issue 9, September 2007, Pages 1029–1036